\chapter{The KATRIN Experiment - An Overview}
\label{ch:katrin}
At the \acs{katrin} experiment, the electron antineutrino mass is inferred from the shape of the beta decay spectrum of tritium. The contribution of the mass on the shape is most dominant in the endpoint region, see also Introduction. Thus, the \acs{katrin} experiment is designed to provide a high-energy resolution, as well as a high event rate in this region of interest. These requirements are met by using a high luminous tritium source and a high-precision spectrometer.
The experimental setup is \SI{70}{\meter} long and can be divided into two parts: the \ac{sts} and the \ac{sds}, see figure~\ref{fig:Chapter2_KATRIN/katrin_bild.jpg}. Neutral tritium gas is injected into the center of the \ac{wgts} and is pumped out at both sides of the source. The \ac{dps} and the \ac{cps} are connected to the source and reduce the tritium density to a bare minimum. Thus, the amount of tritium entering the detector section is negligible. The electrons from the beta decay of tritium are guided adiabatically by a strong magnetic field to the \acs{sds}. There, both the \ac{ps} and the \ac{ms} filter the electrons according to their energy. The electrons are then counted by the detector providing an integral spectrum. The components will be discussed in the following to provide an overview of the \acs{katrin} setup.
\xfigure{tb}{1.}{Chapter2_KATRIN/katrin_bild.jpg}{KATRIN beam line and its components.}{Neutral tritium gas is injected in the \acs{wgts} and pumped out in the \acs{dps} and the \acs{cps}. Beta electrons are guided adiabatically by strong magnetic fields to the spectrometers, where they are sorted by their energy. The detector counts the incoming electrons. The rear section closes off the setup and is equipped with additional monitoring devices. Figure from \cite{PhDHeizmann2019}.}
\section{Windowless Gaseous Tritium Source}
\label{ssec:wgts}
The windowless gaseous tritium source provides beta electrons for the measurement with an activity of up to \SI{e11}{Bq} at highly stable conditions \cite{Aker.352021}. The source consists of a \SI{10}{m} long stainless steel tube that is connected at both sides to turbo-molecular pumps, see figure \ref{fig:Chapter2_KATRIN/wgts.pdf}. Tritium gas is injected into the center of the tube through a specially designed injection chamber. The gas then streams to both sides, where it is removed by the turbo-molecular pumps. It is then processed, filtered and injected again in the center of the source. In this cycle, approximately \SI{40}{g} of tritium is used to maintain a constant flow of gas in the source \citep{Aker.352021}. The actual amount of tritium inside the source can be quantified by the column density. It is equal to the integral of the particle density along the source tube. At nominal source conditions, the column density is \SI{5E17}{molecules\per\centi\meter\squared} \cite{Aker.352021}. For more information on the gas flow, refer to section~\ref{ssec:neutral_gas_flow}.
The \acs{wgts} beam tube is surrounded by a two-phase cryostat, ensuring a constant temperature inside the cavity. The temperature can be adjusted at low temperatures from \SIrange{28}{37}{K} and at higher temperatures from \SIrange{80}{115}{K} \citep{Aker.352021}. Commissioning measurements have shown, that the temperature stays constant over a long period of time, exceeding the design value of \SI{0.1}{\percent\per\hour}. Also, the temperature variation along the beam tube exceeds the design value of \SI{0.1}{\percent} \citep{Aker.352021}. For more information on the temperature profile along the complete source refer to section \ref{sssec:temperature_profile}.
The cryostat also houses superconducting magnets. These magnets provide the field necessary to guide the beta electrons to the detector. The corresponding field is described further in section~\ref{ssec:katrin:B_bg}.
\xfigure{tb}{1.}{Chapter2_KATRIN/wgts.pdf}{Windowless Gaseous Tritium Source.}{Tritium is injected at the center and diffuses to both sides and is pumped out by turbomolecular pumps. A density gradient builds up, depicted in turquoise. The surrounding cryostat ensures stable temperatures and houses superconducting magnets. Figure adapted from \citep{PhDHeizmann2019}.}
\paragraph{Injection Chamber}
Tritium gas is injected in the center of the \acs{wgts} through a specially designed injection chamber, see figure \ref{fig:Chapter2_KATRIN/2D_3D_Inlet_mod1.png}. The gas is streaming through \num{415} orifices, which are designed to \enquote{guarantee a shock-wave free injection without turbulence} \cite{PhDHeizmann2019}. A pressure-controlled buffer vessel ensures a constant injection pressure. Commissioning measurement showed that the pressure is stable at a level of \SI{0.038}{\percent}. This value exceeds the design value of \SI{0.1}{\percent} significantly \cite{Aker.352021}.
\xfigure{h}{0.5}{Chapter2_KATRIN/2D_3D_Inlet_mod1.png}{Injection Chamber.}{Tritium is fed through \num{415} orifices, and streams to both sides of the \acs{wgts}. Figure from \cite{PhDHeizmann2019}.}
The pressure of the vessel is adjusted corresponding to the target density in the source. The inlet pressure $p_{\text{in}}$ directly corresponds to the density right in front of the inlet $n_\text{c}$ and is calculated through the ideal gas law
\begin{equation}
n_\text{in} = \frac{p_\text{in}}{k_b T} \comma
\end{equation}
where $T$ denotes the temperature and $k_b$ Boltzmann's constant. The inlet pressure can be adjusted to produce the needed density inside the source. The pressure for a column density of \SI{5E17}{\mole\per\square\centi\meter} at \SI{30}{\kelvin} is calculated to be \SI{0.337}{Pa}. This value changes for lower column densities or higher temperatures. If, for example, the desired column density is \SI{75}{\percent} of the nominal value at a temperature of \SI{80}{K}, then the pressure needs to be changed to a value of \SI{0.7}{Pa}. For more information on the density along the whole source tube see section \ref{sssec:density_distribution}.
%\paragraph{Source Plasma}
% Some beta electrons hit the surrounding neutral tritium gas, partly ionizing the gas. Hence, more electrons and ions reside inside the \acs{wgts} than from beta decay alone. The total density of these charged particles is closely linked to the cross section of the interactions and the density of the neutral tritium gas. Calculations from Kuckert \cite{PhDKuckert2016} show that the density is approximately \SI{E6}{\per\cubic\centi\meter}. This density is high enough that the charged particles cannot be described as non-interacting particles but as plasma. The source plasma can have a direct influence on the neutrino mass measurement, because it can de- or accelerate electrons. This plasma is investigated in detail in the following chapters.
\section{Differential Pumping Section}
The differential pumping section connects the \acs{wgts} to the cryogenic pumping system, see figure \ref{fig:Chapter2_KATRIN/katrin_bild.jpg}. It acts as an additional pumping stage to reduce the neutral particle flux in the direction of the detectors. Commissioning measurements have shown that the tritium density is reduced in the \acs{dps} by a factor of \num{9.6+-1.0E7} at nominal column density \cite{Aker.352021}. A reduction in tritium flow this large is achieved by the usage of five beam elements, attached to turbo-molecular pumps. These beam tube elements are arranged in a chicane to increase the number of wall collisions of the molecules. Thus, this special arrangement increases the pumping probability.
\xfigure{tb}{0.3}{Chapter2_KATRIN/VacuumpaperDPSfigure_v1.pdf}{Differential Pumping Section.}{Residual gas from the source is pumped out through turbomolecular pumps (yellow). The beam tube is bent in a chicane to maximize wall collisions of neutral gas. The beam tube is surrounded by superconducting magnets (light blue) to transport the beta electrons to the spectrometer section. Ions are removed by dipole electrodes. Figure adapted from \cite{PhDHeizmann2019}.}
The beam tube elements are surrounded by superconducting magnets, which guide electrons through the chicane without collision with the tube walls. This magnetic field would in principle enable ions to traverse to the detector section as well. These ions would act as an additional background to the neutrino mass measurement. Therefore, additional dipole electrodes were installed at the beam tube elements. %They are used in tandem with a single electrode at the end of the \acs{dps} to prevent the movement of ions to the detector section and remove the ions from the source.
The dipole electrodes provide an electric field which is perpendicular to the magnetic field. This configuration induces a $E \times B$ drift in the charged particle movement. The drift distance is dependent on the time spend in the cross field. Beta electrons spend only a short time in the \acs{dps}. Thus, they leave the \acs{dps} almost unhindered. The ions in turn are much slower. Hence, they drift to the beam tube walls, where they neutralize. An additional electrode at the end of the \acs{dps} is set on positive potential. Thus, any remaining ions are reflected back to the \acs{wgts}, which increases the time spent in the cross field even further. In total, ions are hindered from moving towards the detector, while electrons can pass through the \acs{dps}.
\section{Cryogenic Pumping Section}
The cryogenic pumping section concludes the source and transport section and eliminates almost all tritium gas leaving the \acs{dps}. At this stage, mechanical pumping becomes ineffective because of back-diffusion in the pumps. In the \acs{cps} the pumping is achieved by cryosorption. The tube walls are cooled down to \SI{3}{\kelvin} \citep{Aker.352021}. This low temperature builds a cold trap for the tritium gas. The sticking coefficient is increased by the usage of an argon frost coating on a gold-plated surface. The coating must be regenerated after \num{60} days to eliminate the excess tritium. Similar to the \acs{dps} the \acs{cps} beam tube elements are arranged in a chicane to increase the probability that a tritium molecule hits the tube walls. Likewise, superconducting magnets are used to guide the electrons through the setup. See figure \ref{fig:Chapter2_KATRIN/CPS.png} for a graphical representation of the \acs{cps}.
\xfigure{tb}{0.4}{Chapter2_KATRIN/CPS.png}{Cryogenic Pumping Section.}{Residual gas from the \acs{dps} is cryosorped to the \SI{3}{K} cold gold plated beam tube. Additional argon frost enlarges the sticking coefficient. The beam tube is bent in a chicane to maximize wall collisions of neutral gas. The beam tube is surrounded by superconducting magnets (dark red) to transport the beta electrons to the spectrometer section. Figure from \cite{PhDHeizmann2019}.}
Commissioning measurements have shown that the tritium flow rate is reduced at least by a factor of \num{E8}. The sensitivity of this measurement was limited by the sensitivity of the residual gas analyzer. The true reduction factor might even be lower than this value \cite{Aker.352021}. In combination with the \acs{dps} the total reduction factor of the transport section is at least \num{e14}.
The \acs{cps} houses two additional calibration systems, the forward beam monitor \acs{fbm} and a condensed krypton source. Both can be inserted in the flux tube between the last two beam tube elements, see figure \ref{fig:Chapter2_KATRIN/CPS.png}. The \acs{fbm} can measure the relative electron flux originating from the source. Two independent motion systems can move the detector throughout the cross section of the flux tube. Thus, the \acs{fbm} can be used for activity monitoring of the source, as well as monitoring of the magnetic flux tube. The condensed krypton source emits conversion electrons with a small line width. These electrons are then measured by the detector. The condensed krypton source can be used to calibrate the detector section. It can provide \enquote{precise values for the magnetic fields and electric potentials in the analyzing plane, which can be used to cross-check values obtained by simulation...}\cite{Aker.352021}.
\section{Pre- and Main-Spectrometer}
The pre- and main-spectrometer are used to filter the incoming electron flux corresponding to its energy. The spectrometers are used in tandem, where the pre-spectrometer filters out the main portion of low-energy electrons and the main spectrometer is used for high precision filtering. Both spectrometers work according to the same principle: magnetic adiabatic collimation with electrostatic retardation (\acs{mace}). A schematic representation of this principle shown in figure \ref{fig:Chapter2_KATRIN/KATRIN/MACE_adapted.pdf}, and is discussed in the following.
\xfigure{tb}{0.99}{Chapter2_KATRIN/KATRIN/MACE_adapted.pdf}{MAC-E filter principle.}{Beta electrons gyrate around the magnetic field lines, shown in red. The adiabatic reduction of the magnetic field strength converts the perpendicular kinetic energy to parallel kinetic energy. Thus, the momentum vector is oriented in the direction of movement. The electrodes of the spectrometer produce an opposing electric field. Only electrons with an energy higher than $q E$ can pass, and are collected at the detector. Figure adapted from \cite{PhDWandkowsky2013}.}
The strong magnetic field of the \acs{katrin} beam line guides electrons to the spectrometers. The movement of the electrons can be described by a circular motion around a guiding center, which traverses in longitudinal direction. The total kinetic energy is therefore shared by the kinetic energy of the guiding center $E_\parallel$ and by the kinetic energy of the gyro motion $E_\perp$. Beta electrons are emitted isotropically in the source. Thus, the distribution of energy is divided evenly between the two kinetic energy contributions. The relation between the two kinetic energies can be changed by reducing the magnetic field adiabatically.
The magnetic moment $\mu$ is conserved in adiabatic changes of the magnetic field. The magnetic moment calculates to \cite{Aker.1132021}
\begin{equation}
\mu = \frac{E_\perp}{B} = \text{const.} \comma
\end{equation}
where $B$ the magnetic field strength. Hence, a reduction of the magnetic field strength reduces the energy of the cyclotron motion. The energy is transferred to the kinetic energy of the guiding center due to energy conservation. So, the perpendicular kinetic energy is at its lowest at the plane of the lowest magnetic field strength. By applying an electrostatic potential $U_\text{max}$ at this plane position, only electrons with a parallel kinetic energy greater than the retarding potential can pass. Hence, this plane is also called analyzing plane.
The relation between maximal and minimal magnetic field strength ($B_\text{max}$, $B_\text{min}$) defines the minimal energy $\Updelta E$ which could not be converted into parallel direction
\begin{equation}
\frac{\Updelta E}{E} = \frac{B_\text{min}}{B_\text{max}} \fullstop
\end{equation}
The design minimum magnetic field strength of the \acs{ms} is $B_\text{min} = \SI{0.3}{mT}$ and the design maximal magnetic field strength is $B_\text{max} = \SI{6}{T}$ \citep{KATRIN2005}. With a maximal electron energy of the beta decay $E_{max} = \SI{18.6}{keV}$ the energy resolution of the \acs{katrin} spectrometer calculates to $\Updelta E = \SI{0.93}{eV}$ \citep{Aker.352021}.
The reduction of the magnetic field strength also induces an expansion of the magnetic flux tube. To prevent electrons from hitting the walls of the experiment, the flux $\Phi$ needs to be conserved. For an area $A$ perpendicular to the magnetic field $B$ it calculates to
\begin{equation}
\Phi = A \cdot B \fullstop
\end{equation}
In the source section, the design field strength has a value of $B_\text{S} = \SI{3.6}{T}$ \citep{KATRIN2005}. The radius of the source tube is $R = \SI{4.5}{cm}$. Combining this with the minimal magnetic field at the analyzing plane $B_\text{min}$ results in a flux tube diameter in the \acs{ms} of approximately $\SI{10}{m}$.
\section{Detector}
The electron beam passing the analysis plane is collimated by a strong magnet at the end of the \acs{ms}. The individual electrons are then detected by a 148 pixel silicon detector. The electrons are accelerated before detection by an additional electrode, which is kept at \SI{10}{keV}. This procedure reduces backscattering effects and shifts \enquote{the signal peak into a region of lower intrinsic background} \cite{Aker.1132021}. Overall, the detector has a detection efficiency greater than $\epsilon_{\text{det},i} \geq 90\%$ for each individual pixel $i$ \cite{Amsbaugh2015}.
The count rate at the detector is measured for each retarding potential of the \acs{ms}. This integral spectrum of the electron flux of the source is then used to determine the neutrino mass. The count rate $\dot N_i$ for each pixel $i$ is calculated by
\begin{equation}
\dot N_i(U) = \frac{1}{2} \epsilon_{\text{det},i} \cdot \int_{eU}^\infty \frac{\mathrm{d}\Gamma}{\mathrm{d}E}(E,m_\nu^2,E_0)\cdot R_{i}(\vec{x},E,U) \mathrm{d}E \comma
\end{equation}
where $U$ is the potential at the analyzing plane, $R_i$ is the response function, $\frac{\mathrm{d}\Gamma}{\mathrm{d}E}(E,m_\nu^2,E_0)$ is the Tritium beta decay spectrum, see equation~\ref{eq:Neutrino-TdecayRate}, with the endpoint energy $E_0$ and the neutrino mass $m_\nu$ \citep{arXivKleesiek2018}. The response function evaluates all effects where electrons have gained or lost energy inside the \acs{katrin} setup depending on their start position $x$ and weighs them with the probability of transmission at the spectrometers. The response function also includes additional corrections of the Doppler effect or start potential differences of the electrons. The neutrino mass is then determined by an evaluation of the difference between the measured rate and the simulated rate.
\section{Rear Section}
The rear section \acs{rs} is directly connected to the \acs{wgts}, see figure \ref{fig:Chapter2_KATRIN/katrin_bild.jpg}. It closes the beam line in the opposite direction of the detector. All beta electrons, which were reflected by the potential of the analyzing plane, are terminated here at the so-called rear wall. Additionally, the rear section is equipped with various diagnostic systems. These systems include a high resolution angular-selective electron gun and a beta a beta-induced X-ray spectroscopy system (\acs{bixs}).
\paragraph{Rear Wall}
The rear wall is a gold-coated stainless-steel disk with a radius of \SI{7.62}{cm} \cite{PhDKuckert2016}. It is positioned at the end of the beam tube, see figure \ref{fig:Chapter2_KATRIN/KATRIN/figures_rw_1.png}. It is surrounded by a superconducting magnet. This magnet ensures that the magnetic flux tube of the \acs{wgts} is mapped completely to the rear wall. So in general, all charged particles leaving the \acs{wgts} in rear direction will hit the rear wall at some point. The induced current can be measured by an ammeter.
The rear wall naturally exerts a contact potential due to workfunction differences between the rear wall and the surrounding stainless steel of the beam tube. The potential of the rear wall can be adjusted by an additional power supply. This potential is an important boundary condition for the source plasma. Thus, the adjustment of the potential can be used to investigate the plasma.
The rear wall can be illuminated by a highly luminous UV light source \cite{PhDFulst2020}. The UV light produces photo electrons, which move into the source. These electrons can either be used for commissioning measurements of the empty source or they can be used as an additional source of negative charges in the source plasma, see section~\ref{sec:measurements_on_the_source_plasma}.
\xfigure{tb}{1.0}{Chapter2_KATRIN/KATRIN/figures_rw_1.png}{Rear wall chamber.}{(1),(4) Mounting flanges, (2) rear wall, (3) central hole for \acs{egun} beam transmission. Figure taken from \cite{Aker.352021}.}
%\newpage
\paragraph{Electron Gun}
The \ac{egun} is situated behind the rear wall. It produces high-energy electrons up to \SI{30}{keV} with an intrinsic energy spread of approximately \SI{0.2}{eV} \cite{PhDBabutzka2014}. These electrons are fed through a hole with a diameter of \SI{5}{mm} inside the rear wall. The electron beam can be directed through magnetic and electric dipoles to cover the full beam tube cross section.
\enquote{The e-gun allows studying electromagnetic characteristics along the main beam line and has been used in the commissioning of various beam line sections. It is also suited to investigate and monitor source characteristics, such as electron scattering and the column density stability} \cite{Aker.352021}.
\paragraph{BIXS}
The \acs{bixs} system is used to measure the activity of the source during neutrino mass measurement runs. It is located at one of the access pumps of the rear wall chamber. The \acs{bixs} system consists of a silicon drift detector, which measures the bremsstrahlung of beta electrons hitting the rear wall. The count rate directly corresponds to the activity of the source. Thus, the \acs{bixs} system can also be used to monitor the column density. In commissioning measurements, it was shown that the system allows for a \enquote{stability monitoring on the \SI{0.1}{\percent} level in measurement cycles of duration \SI{100}{s}} \cite{Aker.352021}.